Keysight CXG, EXG, and MXG X-Series Signal Generators Service Guide 31
Overview
After an Instrument Repair
After an Instrument Repair
If any instrument assemblies have been repaired or replaced, perform the
related adjustments and performance verification tests. Most of these tests are
done using the N7822A Calibration Application Software for CXG/EXG/MXG
Signal Generators. Refer to Chapter 16, “Pre and Post-Repair Procedures” for
a list of post-repair adjustments and performance tests based on which
hardware has been changed.
Information regarding the N7822A Calibration Application Software for
CXG/EXG/MXG Signal Generators can be found at
http://www.keysight.com/find/calibrationsoftware