EasyManuals Logo

Keysight Technologies N5183B MXG Service Guide

Keysight Technologies N5183B MXG
451 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #384 background imageLoading...
Page #384 background image
384 Keysight CXG, EXG, and MXG X-Series Signal Generators Service Guide
Assembly Replacement
A7A1 Real Time Aux Power Supply
A7A1 Real Time Aux Power Supply
Tools Required
T-10 driver
needle-nose pliers
Removal Procedure
Refer to Figure 15-31 for this procedure.
1. Disconnect the power cord.
2. Remove the covers from the signal generator. Refer to “Outer Cover” on
page 336 and “Inner Top Cover” on page 338.
3. Place the signal generator flat with the A7 Real Time BBG assembly facing
up and the rear panel facing you.
4. Refer to Figure 15-31. Remove power cable W12 from the A7A1 Real Time
Aux Power Supply (1).
5. Remove power cable W10 from A7A1 Real Time Aux Power Supply
assembly.
6. Refer to Figure 15-32. Remove the four screws (1) attaching the A7A1
Real Time Aux Power Supply cover. Remove cover.
7. Refer to Figure 15-33. Remove the fours screws (2) attaching the A7A1
Real Time Aux Power Supply to the chassis.
8. Remove the A7A1 Real Time Aux Power Supply.
Replacement Procedure
1. Reverse the order of the removal procedure.
2. Recommended torque for all T-10 screws is 9 in–lbs.
3. Perform the post-repair procedures that pertain to this replacement
procedure. See Chapter 16, “Pre and Post-Repair Procedures.”

Table of Contents

Other manuals for Keysight Technologies N5183B MXG

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies N5183B MXG and is the answer not in the manual?

Keysight Technologies N5183B MXG Specifications

General IconGeneral
BrandKeysight Technologies
ModelN5183B MXG
CategoryPortable Generator
LanguageEnglish

Related product manuals