EasyManuals Logo

Keysight Technologies N5183B MXG Service Guide

Keysight Technologies N5183B MXG
451 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #403 background imageLoading...
Page #403 background image
Keysight CXG, EXG, and MXG X-Series Signal Generators Service Guide 403
Assembly Replacement
Rear Panel (N5171B/81B)
Rear Panel (N5171B/81B)
Tools Required
T-10 driver
5/8” hex-nut driver (p/n: 8710-2546)
Removal Procedure
Refer to Figure 15-41 for this procedure.
1. Disconnect the power cord.
2. Remove the covers from the signal generator. Refer to “Outer Cover” on
page 336, “Inner Top Cover” on page 338, and “Inner Bottom Cover” on
page 340.
3. Place the signal generator flat with the rear panel facing you.
4. If Option 006 is installed, refer to Figure 15-42. Remove the Flash Memory
card (1) and cover (2).
5. Using the 5/8” hex-nut driver, remove the nuts and washers (1) securing
the A3 RF assembly BNC’s to the rear panel.
6. Using the T-10 driver, remove the two screws (2) and the five screws (3)
securing the rear panel to the signal generators chassis.
7. Pull the rear panel away from the signal generator’s chassis.
Replacement Procedure
1. Reverse the order of the removal procedure.
2. Recommended torque for all screws is 9 in–lbs.
3. Recommended torque for all BNC connectors is 21 in–lbs.

Table of Contents

Other manuals for Keysight Technologies N5183B MXG

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies N5183B MXG and is the answer not in the manual?

Keysight Technologies N5183B MXG Specifications

General IconGeneral
BrandKeysight Technologies
ModelN5183B MXG
CategoryPortable Generator
LanguageEnglish

Related product manuals