Pre and Post-Repair Procedures
Additional Tasks
Finalize RF Assembly Installation
This task is for N5166B, N5172B, and N5182B instruments and will require the
use of additional test equipment.
Perform this required task by following this procedure:
1. Press Utility, More, Service to enter the Service menu.
2. Press Install Assembly, Finalize RF Assembly Installation.
3. Connect and configure a supported power meter to the signal source, or a
communication bus that it supports. See Chapter 5, “Service and Utility
Menus,” for detailed information on selecting and configuring this
equipment.
4. Once the required equipment has been connected and verified press
Confirm Installation of RF Assembly.
This procedure could take up to 1 hour or more to complete. Once this
procedure has completed continue with the next task associated with the A3
RF Assembly replacement.
Finalize BBG Assembly Installation
This task is for N5166B, N5172B, and N5182B instruments and will require the
use of additional test equipment.
Perform this required task by following this procedure:
1. Press Utility, More, Service to enter the Service menu.
2. Press Install Assembly, Finalize BBG Assembly Installation.
3. Connect and configure a supported power meter to the signal source, or a
communication bus that it supports. See Chapter 5, “Service and Utility
Menus,” for detailed information on selecting and configuring this
equipment.
4. Once the required equipment has been connected and verified press
Confirm Installation of BBG Assembly.
This procedure could take up to 1 hour or more to complete. Once this
procedure has completed continue with the next task associated with the A2
Vector BBG Assembly replacement.
Restore Licenses
When the licenses for frequency range and other options are installed in the
instrument they are saved in non-volatile memory on the A5 CPU assembly.
When the instrument is originally shipped from the factory a backup copy of
these licenses is also saved on the A3 RF assembly. If the A5 CPU assembly is
replaced the licenses that are backed up on the A3 RF assembly will need to be
restored to the A5 CPU assembly so that the proper functionality is restored to
the instrument.

Table of Contents

Other manuals for Keysight Technologies N5183B MXG

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies N5183B MXG and is the answer not in the manual?

Keysight Technologies N5183B MXG Specifications

General IconGeneral
BrandKeysight Technologies
ModelN5183B MXG
CategoryPortable Generator
LanguageEnglish

Related product manuals