Setting Up the Oscilloscope 1
Keysight Infiniium 90000A Series Oscilloscopes User’s Guide 25
Verifying Basic Oscilloscope Operation
1 Connect one end of the calibration cable using SMA-to-precision BNC
adapters to channel 1.
2 Connect the other end of the calibration cable to the Aux Out connector on the
front panel.
3 Press [Default Setup] on the front panel.
The display will pause momentarily while the oscilloscope is configured to its
default settings.
4 Press [Auto Scale] on the front panel.
The display will pause momentarily while the oscilloscope adjusts the time/div
setting and vertical scale. You should then see a square wave with about four
cycles on screen and a peak-to-peak amplitude of approximately five divisions.
If you do not see the waveform, make sure your power source is adequate, the
oscilloscope is properly powered on, and the probe is connected securely to the
front-panel channel input BNC and to the probe calibration output.
5 Move the mouse around the mouse surface and verify that the on-screen
pointer follows the mouse movement.
6 Touch the pointer of the touch screen stylus to the surface of the screen and
move it around while verifying that the pointer follows the movement.
Figure 10 Verifying basic oscilloscope operation
Aux Out Aux Trig