Using the Oscilloscope 2
Keysight Infiniium 90000A Series Oscilloscopes User’s Guide 47
Forcing a Default Setup
If your Infiniium oscilloscope is not working properly when you start it up, follow
these steps to perform a default setup and return the oscilloscope to normal
operation.
1 Click Control > Default Setup or press the [Default Setup] key.
2 If the oscilloscope is still not working properly, click Control > Factory Default to
return the oscilloscope to the default settings it had when it left the factory.
3 If the oscilloscope is still not working properly, turn it off.
4 Turn the oscilloscope back on. If it does not successfully restart, try recycling
the power again.
5 As soon as the Windows 7 load screen disappears, press [Default Setup]. If the
oscilloscope still does not successfully restart, follow the instructions for
recovering the hard drive.
Infiniium hard drive recovery
Follow these steps to recover your Infiniium hard drive.
1 Turn off the oscilloscope.
2 Connect the keyboard and mouse to the USB host ports on the back of the
oscilloscope.
3 Turn on the oscilloscope.
4 As soon as you see the message press F2 to boot the Infiniium Recovery System,
press [F2] and follow the on-screen instructions.
5 Once the recovery process is finished and the oscilloscope is running, check in
the About Infiniium dialog box under installed options to see if all of the options
you ordered are installed. If the options are not installed, install them using the
license keys provided on the oscilloscope option license certificates you
received, or refer to the back of the oscilloscope.