496 Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
31 MIL-STD-1553/ARINC 429 Triggering and Serial Decode
2 In the Search Menu, press the Search softkey; then, turn the Entry knob to select
the serial slot (Serial 1 or Serial 2) on which the ARINC 429 signal is being
decoded.
3 Press Search; then, select from these options:
• Label — finds the specified label value.
Label values are always displayed in octal.
• Label + Bits — finds the label and the other word fields as specified.
• Parity Error — finds words with a parity error.
• Word Error — finds an intra-word coding error.
• Gap Error — finds an inter-word gap error.
• Word or Gap Error — finds either a Word or Gap Error.
• All Errors — finds any of the above errors.
For more information on searching data, see “Searching Lister Data" on page 152.
For more information on using the [Navigate] key and controls, see “Navigating the
Time Base" on page 75.