36 Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
1 Getting Started
 
The trigger source is selected by looking for the first valid waveform starting with 
external trigger, then continuing with the lowest number analog channel up to the 
highest number analog channel, and finally (if digital probes are connected) the 
highest number digital channel.
During Autoscale, the delay is set to 0.0 seconds, the horizontal time/div (sweep 
speed) setting is a function of the input signal (about 2 periods of the triggered 
signal on the screen), and the triggering mode is set to Edge.
Compensate Passive Probes
Each oscilloscope passive probe must be compensated to match the input 
characteristics of the oscilloscope channel to which it is connected. A poorly 
compensated probe can introduce significant measurement errors.
1 Input the Probe Comp signal (see “Input a Waveform" on page 33).
2 Press [Default Setup] to recall the default oscilloscope setup (see “Recall the 
Default Oscilloscope Setup" on page 33).
3 Press [Auto Scale] to automatically configure the oscilloscope for the Probe 
Comp signal (see “Use Autoscale" on page 34).
4 Press the channel key to which the probe is connected ([1], [2], etc.).
5 In the Channel Menu, press Probe.
6 In the Channel Probe Menu, press Probe Check; then, follow the instructions 
on-screen.
If necessary, use a nonmetallic tool (supplied with the probe) to adjust the 
trimmer capacitor on the probe for the flattest pulse possible.
On N2894A probes, the trimmer capacitor is located on the probe BNC 
connector.