222 Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
13 Acquisition Control
 
• Normal — at slower time/div settings, normal decimation occurs, and there is 
no averaging. Use this mode for most waveforms. See “Normal Acquisition 
Mode" on page 222.
• Peak Detect — at slower time/div settings, the maximum and minimum 
samples in the effective sample period are stored. Use this mode for 
displaying narrow pulses that occur infrequently. See “Peak Detect 
Acquisition Mode" on page 222.
• Averaging — at all time/div settings, the specified number of triggers are 
averaged together. Use this mode for reducing noise and increasing 
resolution of periodic signals without bandwidth or rise time degradation. 
See “Averaging Acquisition Mode" on page 225.
• High Resolution — at slower time/div settings, all samples in the effective 
sample period are averaged and the average value is stored. Use this mode 
for reducing random noise. See “High Resolution Acquisition Mode" on 
page 227.
Normal Acquisition Mode
In Normal mode at slower time/div settings, extra samples are decimated (in other 
words, some are thrown away). This mode yields the best display for most 
waveforms.
Peak Detect Acquisition Mode
In Peak Detect mode at slower time/div settings, minimum and maximum samples 
are kept in order to capture infrequent and narrow events (at the expense of 
exaggerating any noise). This mode displays all pulses that are at least as wide as 
the sample period.
For InfiniiVision 3000T X-Series oscilloscopes, which have a maximum sample rate 
of 5 GSa/s, a sample is taken every 200 ps (sample period).
See Also • “Glitch or Narrow Pulse Capture" on page 223
• “Using Peak Detect Mode to Find a Glitch" on page 224