Repair Guide 6
Keysight N1913/1914A Service Guide 107
Remove the top clamshell (Figure 6-3). Disconnect the mains
power connector from the top clamshell. Disconnect the ribbon
cable from the main board. Disconnect both earth spade
connectors from the top clamshell. Remove top clamshell.
Figure 6-3 Remove the top clamshell
– Remove the PSU safety cover (Figure 6-4). Use the T10 Torx
screwdriver bit to remove the four screws attaching the PSU
safety cover to the top clamshell. Lift and remove the safety
cover.
– Remove the PSU cable guide (Figure 6-4). Use the T10 Torx
screwdriver bit to remove the screw attaching the cable
guide to the top clamshell. Lift and remove cable guide.
Figure 6-4 Remove the PSU safety cover
and cable guide
Instructions Visual