EasyManua.ls Logo

Keysight N1914A - Replacing the Calibrator Semi-Rigid;Split Ferrite; Figure 6-19 Separate the Two Halves of the Ferrite

Keysight N1914A
141 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Repair Guide 6
Keysight N1913/1914A Service Guide 131
Replacing the Calibrator Semi-Rigid/Split Ferrite
Instructions Visual
Separate the two halves of the ferrite (Figure 6-19).
Position the ferrite such that it's furthest edge is 120 mm
(4 ¾”) from the bend of the semi-rigid.
Hold the ferrite in place by applying a coating of silicone or
silicone-rubber compound (e. g. RTV) along that 20 mm (¾”)
section of the semi-rigid.
Join both halves of the ferrite, keeping the mating surfaces
free of the silicone compound if possible.
Figure 6-19 Separate the two halves of the
ferrite

Table of Contents