Specifications and Characteristics 1
Keysight N1913/1914A Service Guide 43
1 mW Power Reference
The 1 mW Power Reference is provided for calibration of the E-Series, 8480
Series, and N8480 Series power sensors.
Power output
1.00 mW (0.0 dBm)
Factory set to ±0.4% traceable to the National Physical
Laboratories (NPL), UK
Accuracy
±1.2% (0 – 55 ºC)
±0.4% (25 ±10 ºC)
Frequency 50 MHz nominal
SWR
1.08 (0 - 55 ºC)
1.05 (typical)
Connector type Type N (f), 50 Ω