6 Repair Guide
110 Keysight N1913/1914A Service Guide
Remove the EMI earth wires (Figure 6-7). Use the 9/32” socket
to remove the Hex nut attaching the EMI earth wires to the
calibrator assembly. Remove the earth wires and washers,
taking note of the assembly order.
Figure 6-7 Remove the EMI earth wires
Disconnect the front panel cable (Figure 6-8). Depress both
sides of the connector holding the ribbon cable to eject it.
Figure 6-8 Disconnect the front panel
cable
Disconnect the semi-rigid cable (Figure 6-9).
Note:
This only applies to Option 003 units. Use the 5/16” spanner to
disconnect the semi-rigid cable from the N-type connector on
the rear panel.
Figure 6-9 Disconnect the semi-rigid cable