Repair Guide 6
Keysight N1913/1914A Service Guide 113
Reassembly Instructions
Instructions Visual
The reassembly process is simply the reverse of the
disassembly process. However, there are various points to be
aware of.
– USB/LAN connectors must rest on top of the rear panels'
EMC spring fingers.
– The position of the cable clamp depends on whether Option
101 or 003 is fitted.
– The main board connector from the PSU must be pushed
firmly to fully engage it.
– Take care not to trap any cables when fitting the top
clamshell.
– Analog recorder output connections (Figure 6-14). Ensure
that recorder 1 is plugged into the rear connector. Where
applicable, recorder 2 is plugged into the connector nearer
to the front.
Figure 6-14 Analog recorder output
connections
Sensor flex connections (Figure 6-15)
– A - Front, Channel A (for Option 101)
– B - Front, Channel B (for Option 101)
– C - Rear, Channel A (for Option 003)
– D - Rear, Channel B (for Option 003)
1 J28 - front Ch 1
2 J127 - front Ch 2
3 J23 - rear Ch 2
4 J18 - rear Ch 1