Repair Guide 6
Keysight N1913/1914A Service Guide 113
Reassembly Instructions
Instructions Visual
The reassembly process is simply the reverse of the
disassembly process. However, there are various points to be
aware of.
USB/LAN connectors must rest on top of the rear panels'
EMC spring fingers.
The position of the cable clamp depends on whether Option
101 or 003 is fitted.
The main board connector from the PSU must be pushed
firmly to fully engage it.
Take care not to trap any cables when fitting the top
clamshell.
Analog recorder output connections (Figure 6-14). Ensure
that recorder 1 is plugged into the rear connector. Where
applicable, recorder 2 is plugged into the connector nearer
to the front.
Figure 6-14 Analog recorder output
connections
Sensor flex connections (Figure 6-15)
A - Front, Channel A (for Option 101)
B - Front, Channel B (for Option 101)
C - Rear, Channel A (for Option 003)
D - Rear, Channel B (for Option 003)
1 J28 - front Ch 1
2 J127 - front Ch 2
3 J23 - rear Ch 2
4 J18 - rear Ch 1