EasyManua.ls Logo

Keysight N1914A - Figure 1-5 U2000 Series Settling Time with Auto-Filter

Keysight N1914A
141 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1 Specifications and Characteristics
40 Keysight N1913/1914A Service Guide
Figure 1-5 U2000 Series settling time with auto-filter
-50 dBm
+10 dBm
+2 dBm
-40 dBm
45 ms
90 ms
2.6 s
2.7 s
460 ms
2.8 s
39 s
42 s
42 s
-4 dBm
-10 dBm
-20 dBm
-30 dBm
Normal
mode
45 ms
82 ms
1.3 s
1.5 s
450 ms
1.6 s
20 s
24 s
24 s
x2
mode
Maximum sensor power
within a range
U2000/1/2/4A
U2000/1/2H U2000/1B
+20 dBm +40 dBm
+12 dBm +32 dBm
+6 dBm +26 dBm
0 dBm
-10 dBm
-20 dBm
-30 dBm
-40 dBm
+20 dBm
+10 dBm
0 dBm
-10 dBm
-20 dBm
Minimum sensor power
within a range
Sensor
dynamic
range
Typical
settling
times
0 dBm
0 dBm
+12 dBm
+10 dBm

Table of Contents