2 Performance Tests
58 Keysight N1913/1914A Service Guide
Figure 2-5 Absolute accuracy test setup
Test Method
1 Connect the equipment as shown in Figure 2-5.
2 Unplug the power cord from the range calibrator. Eliminate ground loops to
enable the 11683A (Option H01) range calibrator to operate properly.
11683A range
calibrator front panel
11683A range
calibrator rear panel
BNC T-joint connector
BNC cable
BNC cable
Banana
connector
Banana
connector
33250A function
generator
3458A digital
multimeter
N1913/1914A EPM
Series power meters
NOTE
The procedure details the key presses required on the Keysight N1913A EPM
Series power meters. For the Keysight N1914A EPM Series power meters, the
equivalent key presses should be performed on both channels.