1 Specifications and Characteristics
24 Keysight N1913/1914A Service Guide
Power Meter Specifications
Frequency Range
9 kHz to 110 GHz, power sensor dependent
Power Range
–70 dBm to +44 dBm (100 pW to 25 W), power sensor dependent
Power Sensors Compatibility
Keysight 8480 Series power sensors
Keysight E9300 E-Series average power sensors
Keysight E4410 E-Series average power sensors
Keysight N8480 Series power sensors
Keysight U2000 Series average USB power sensors
Single Sensor Dynamic Range
90 dB maximum (Keysight E-Series power sensors)
50 dB maximum (Keysight 8480 Series power sensors)
55 dB maximum (Keysight N8480 Series power sensors)
80 dB maximum (Keysight U2000 Series USB power sensors)
Display Units
Absolute: Watts (W) or dBm
Relative: Percent (%) or dB
Display Resolution
Selectable resolution of: 1.0, 0.1, 0.01, and 0.001 dB in logarithmic mode; or 1, 2,
3, and 4 significant digits in linear mode
Default Resolution
0.01 dB in logarithmic mode or three digits in linear mode