6 Repair Guide
118 Keysight N1913/1914A Service Guide
Front panel assembly (Front connectors option)
Remove handle, bumpers, and top clamshell (including PSU)
Disconnect the sensor RF connections from the main board
Disconnect the sensor flex connection(s) from the main
board
Disconnect the calibrator assembly cable connection from
the main board
Disconnect the EMI earth wires from the calibrator assembly
Disconnect main board ribbon cable from the front panel
assembly
Front panel assembly can now be removed
Front panel assembly
(Rear connectors option)
Remove handle, bumpers, and top clamshell (including PSU)
Disconnect calibrator semi-rigid from the rear panel
assembly
Disconnect the calibrator assembly cable connection from
the main board
Disconnect the EMI earth wires from the calibrator assembly
Disconnect main board ribbon cable from the front panel
Front panel assembly can now be removed
Sensor flex assembly
(Front connectors option)
Remove front panel assembly as previously described
Use the N1912-61807 special tooling kit to remove the
sensor flex assembly
Sensor flex assembly
(Rear connectors option)
Remove handle and front/rear bumpers
Remove top clamshell (including PSU)
Disconnect the sensor RF connections from the main board
Disconnect the sensor flex connection from the main board
Use the N1912-61807 special tooling kit to remove the
sensor flex assembly
Disassembly of replacement part Instructions