105
Command Reference
DISPlay Subsystem
Display Line On/Off
:DISPlay:WINDow:TRACe:Y:DLINe:STATe OFF|ON|0|1
:DISPlay:WINDow:TRACe:Y:DLINe:STATe?
Turns the display line on or off.
*RST:
Off
Key access: > Display Line On Off
Normalized Reference Level
:DISPlay:WINDow:TRACe:Y[:SCALe]:NRLevel <rel_ampl>
:DISPlay:WINDow:TRACe:Y[:SCALe]:NRLevel?
Sets the normalized reference level. See command
:CALCulate:NTData[STATe] OFF|ON|0|1
*RST: 0 dB
Range: –327.6 to 327.6 dB
Key access:
> Tracking Generator > More > Normalize > Norm Ref Lvl
Normalized Reference Level Position
:DISPlay:WINDow:TRACe:Y[:SCALe]:NRPosition <integer>
:DISPlay:WINDow:TRACe:Y[:SCALe]:NRPosition?
Selects the position of the normalized reference level. The top and
bottom graticule lines correspond to 10 and 0, respectively.
See
command
:CALCulate:NTData[STATe] OFF|ON|0|1
*RST: 10
Range: integer
Key access:
> Tracking Generator > More > Normalize > Norm Ref Posn