EasyManua.ls Logo

Keysight N9322C - Instrument Subsystem

Keysight N9322C
223 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
69
Command Reference
Instrument Subsystem
Instrument Subsystem
The instrument subsystem includes commands selecting the
instrument modes or power measurement mode.
Instrument Mode Switch
:INSTrument[:SELect] SA|TGENerator|MA|POWme-
ter|CAT
:INSTrument[:SELect]?
This command selects the instrument mode of N9322C. All
the instrument modes are under Mode list of N9322C unit.
SA - Spectrum Analyzer mode.
TGENerator - Tracking Generator mode.
MA - Modulation Analysis mode.
POWmeter - Power Meter mode.
CAT - Reflection Measurement mode
*RST: SA
Example: :INSTrument TGENerator
Key access: MODE
Measurement Mode Switch
:INSTrument:MEASure OFF|CHPower|ACPR|OBW|SPEC-
trogram|SEM|CHScanner
:INSTrument:MEASure?
This command selects the specific power measurement mode.
CHPower - Channel Power Measurement
ACPR - Adjacent Channel Power Ratio Measurement
OBW - Occupied Bandwidth Measurement
SPECtrogram - Spectrum Monitor
SEM - Spectrum Emission Mask

Table of Contents

Related product manuals