Command Reference
System Subsystem
54
System Subsystem
This subsystem is used to set the controls and parameters
associated with the overall system communication. These
functions are not related to instrument performance.
Screen Brightness Setting
:SYSTem:SCReen:BRIGhtness <value>
:SYSTem:SCReen:BRIGhtness?
This command sets the screen brightness of the instrument.
Range: 0 to 10
Example: :SYSTem:SCReen:BRIGhtness 4
Key access: System> Screen Setting
Auto Screen Brightness State
:SYSTem:SCReen:BRIGhtness:AUTO OFF|ON|0|1
:SYSTem:SCReen:BRIGhtness:AUTO?
This command turns on/off auto screen brightness state.
Example: :SYSTem:SCReen:BRIGhtness:AUTO ON
Key access: System> Screen Setting
Set System Time
:SYSTem:TIME <“hhmmss”>
:SYSTem:TIME?
Sets the system time of the instrument.
Hour must be an integer 0 to 23.
Minute must be an integer 0 to 59.
Second must be an integer 0 to 59.
Example: :SYSTem:TIME “150233”
Key access: System> Time/Date > Time