Initiate Subsystem
Initiate Subsystem
The INITiate subsystem is used to control the initiation of
the spectrum analyzer sweep operation.
Continuous or Single Sweep
:INITiate:CONTinuous OFF|ON|0|1
:INITiate:CONTinuous?
Selects whether the sweep operation is continuously initiated
or not.
This command affects sweep if not in a measurement, and
affects trigger when in a measurement. This corresponds to
continuous sweep or single sweep operation when not in a
measurement, and continuous measurement or single
measurement operation when in a measurement.
When not in a measurement, this command does the
following:
When ON at the completion of each sweep cycle, the
sweep system immediately initiates another sweep cycle.
When OFF, the sweep system remains in an “idle” state
until CONTinuous is ON or :INITiate[:IMMediate] is
received. On receiving the :INITiate[:IMMediate]
command, it will go through a single sweep cycle, and
then return to the “idle” state.
The query returns 1 or 0 into the output buffer. 1 is
returned when there is continuous sweeping. 0 is
returned when there is only a single sweep.
When in a measurement, this command does the following:
When ON at the completion of each trigger cycle, the
trigger system immediately initiates another trigger cycle.
When OFF, the trigger system remains in an “idle” state
until CONTinuous is ON or :INITiate[:IMMediate] is
received. On receiving the :INITiate[:IMMediate]
command, it will go through a single trigger cycle, and
then return to the “idle” state.