EasyManuals Logo

Keysight N9322C User Manual

Keysight N9322C
223 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #83 background imageLoading...
Page #83 background image
75
Command Reference
Sense Subsystem
Amplitude Subsection
Reference level
:DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel: <value>
:DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel?
This command sets the reference level for the Y- axis.
*RST: 0 dB
Range: –140.00 to +20.00 dBm
Example: :DISPlay:WINDow:TRACe:Y:RLEVel 10dBm
Key access: Amptd > Ref Level
Attenuation
[:SENSe]:POWer[:RF]:ATTenuation <value>
[:SENSe]:POWer[:RF]:ATTenuation?
Set the input attenuator of the spectrum analyzer.
Range: 0 to 50 dB
Example: :POWer:ATTenuation 30dB
Key access: Amptd > Attenuation
Input Port Attenuator Auto State
[:SENSe]:POWer[:RF]:ATTenuation:AUTO OFF|ON|0|1
[:SENSe]:POWer[:RF]:ATTenuation:AUTO?
This command turns on/off auto input port attenuator state.
*RST: On
Example: :POWer:ATTenuation:AUTO?
Key access: Amptd > Attenuation > Auto

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight N9322C and is the answer not in the manual?

Keysight N9322C Specifications

General IconGeneral
BrandKeysight
ModelN9322C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals