EasyManua.ls Logo

Keysight N9322C - Tracking Generator Option Subsystem

Keysight N9322C
223 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command Reference
Tracking Generator Option Subsystem
212
Tracking Generator Option Subsystem
The Tracking Generator Option Subsystem provides you the
SCPI command reference for the Tracking Generator
funciton.Use the command :INSTrument TGENerator to
access this subsystem.
Option TG7 is required to enable this function.
Tracking Generator Output Amplitude
:SOURce:POWer[:LEVel][:IMMediate][:AMPLitude] <value>
:SOURce:POWer[:LEVel][:IMMediate][:AMPLitude]?
*RST: –20 dBm
Range: 0 to –30 dBm
Example: :SOURce:POWer -10 dBm
Key access: Meas > Amplitude
Tracking Generator Amplitude On/Off
:OUTPut[:STATe] OFF|ON|0|1
:OUTPut[:STATe]?
Use this command to turn/off the tracking generator output.
*RST: Off
Key access: Meas > Amplitude On Off
Tracking Generator Amplitude Offset
:SOURce:CORRection:OFFSet <value>
:SOURce:CORRection:OFFSet?
This command sets the amplitude offset of the tracking
generator output. This command is valid after tracking
generator is enabled.
*RST: 0 dB
Range: 0 to –30 dBm
Example: :SOURce:CORRection:OFFSet -10 dB
Key access: Meas > Amptd Offset