211
Command Reference
Power Meter Option Subsystem
This command defines trigger delay time of power meter
measurement.
Range: 0 to 1 s
Example: :TRIGger:SEQuence:DELay 1
Key access: Meas > Peak Setup > Trig/Acq > Delay
Power Meter Trigger Slope Mode
:TRIGger[:SEQuence]:SLOPe POSitive|NEGative
:TRIGger[:SEQuence]:SLOPe?
This command toggles trigger slope mode between rise and
fall.
Example: :TRIGger:SEQuence:SLOPe POSitive
Key access: Meas > Peak Setup > Trig/Acq > More 1 of 2 > Slop
Power Meter Trigger Holdoff Time
TRIGger[:SEQuence]:HOLDoff <value>
TRIGger[:SEQuence]:HOLDoff?
This command defines the trigger hold off time of power
meter.
Range: 1 us to 400 ms
Example: :TRIGger:SEQuence:HOLDoff 0.1
Key access: Meas > Peak Setup > Trig/Acq > More 1 of 2 > Holdoff
Power Meter Trigger Hysteresis Value
:TRIGger[:SEQuence]:HYSTeresis <value>
:TRIGger[:SEQuence]:HYSTeresis?
This command defines the trigger hysteresis value of power
meter.
Example: :TRIGger:SEQuence:HYSTeresis 1
Key access: Meas > Peak Setup > Trig/Acq > More 1 of 2 > Hysteresis