Command Reference
Power Meter Option Subsystem
210
Set Trigger Source
:TRIGger1:SOURce EXTernal|INTernal
:TRIGger1:SOURce?
Use this command to toggle the trigger source between
external and internal. It’s only available for continuous
trigger or single trigger.
Example: :TRIGger1:SOURce EXTernal
Key access: Meas > Peak Setup > Trig/Acq > Source
Power Meter Auto Trigger Level State
:TRIGger[:SEQuence]:LEVel:AUTO OFF|ON|0|1
:TRIGger[:SEQuence]:LEVel:AUTO?
This command turns on/off auto trigger level state of power
meter measurement.
Example: :TRIGger:SEQuence:LEVel:AUTO 1
Key access: Meas > Peak Setup > Trig/Acq > Mode
Set Trigger Level
:TRIGger[:SEQuence]:LEVel <value>
:TRIGger[:SEQuence]:LEVel?
This command defines the trigger level of power sensor. It’s
only available for continuous trigger or single trigger. The
trigger source must be internal, and auto trigger level state
must be off (set to normal mode).
Range: –100 to 0 dBm
Example: :TRIGger:SEQuence:LEVel -20dBm
Key access: Meas > Peak Setup > Trig/Acq > Level
Power Meter Trigger Delay Time
:TRIGger[:SEQuence]:DELay <value>
:TRIGger[:SEQuence]:DELay?