79
Command Reference
Sense Subsystem
Current Correction Select
[:SENSe]:CORRection:SELect COR1|COR2|COR3|COR4
[:SENSe]:CORRection:SELect?
Set current correction for load COR file onto proper
CorrectionX.
Example: :CORRection:SELect?
Input Impedance selection
[:SENSe]:CORRection:IMPedance[:INPut][:MAGNitude]
OHM50|OHM75
[:SENSe]:CORRection:IMPedance[:INPut][:MAGNitude]?
Toggles the input impedance correction between 50 Ω and
75 Ω. in spectrum analyzer.
*RST: OHM50
Example: :CORRection:IMPedance OHM75
Key access: Amptd > More > Impedance
PSD Mode State
:DISPlay:WINDow:TRACe:Y:PSD OFF|ON|0|1
:DISPlay:WINDow:TRACe:Y:PSD?
Use this command to turn on/off the power spectral density
display mode.
Example: :DISPlay:WINDow:TRACe:Y:PSD 1
Key access: Amptd > More > PSD Function > On/Off