Chapter 2 59
Test Descriptions
Noise Figure Range and Accuracy
6. Set the frequency to 50 MHz (press the Frequency/Points key and Fixed 
Freq
, 5, 0, MHz).
7. Set the averaging to 101 (press the 
Averaging/Bandwidth key and 
Average Mode (Point), Averages, 1, 0, 1, Enter, Averaging On).
8. Select the display to meter mode (press the 
Format key and Format, 
Meter).
9. Fix the IF Gain (press the 
System key then More, More and Service. 
Press 
Enter Password, then -2010 and Enter. Then press Service , IF 
Test
 and Fix IF Gain On. Set the IF Gain Value to 16 and press IF Gain 
Value, 1, 6
 and Enter.
Step 5. Restart the NFA measurement sweep by pressing the 
Restart key.
Step 6. Record the measured P
hot
 value in the Worksheet as the -54dB Ref 1 
Level.
Step 7. Step the attenuator in 1dB steps from 1 - 11dB recording the measured 
P
hot
 at each point in the Worksheet as the -55 to -65dB levels. Remember 
to press the 
Restart key at each measurement.
Step 8. Set the synthesizer power level to -65dBm (press the 
Power Level key, 
then 
-, 6, 5, dB(m)). 
Step 9. Reset the attenuator to 0dB and press the 
Restart key to restart the NFA 
sweep.
Step 10. Record the measured P
hot
 value in the Worksheet as the -65dB Ref 2 
Level.
Step 11. Step the attenuator in 1dB steps from 1 - 11dB recording the measured 
Phot at each point in the Worksheet as the -66 to -76dB levels. 
Remember to press the 
Restart key at each measurement.
Step 12. Reset the IF Gain (press the 
System key then More, More and Service. 
Press 
Enter Password, then -2010 and Enter. Then press Service , IF Test 
and 
Fix IF Gain Off).
Step 13. Enter the attenuator metrology data information in the Worksheet 
provided as Actual Attenuation using equation 1.
Step 14. Calculate the Measured Attenuation and enter the values in the 
Worksheet using equation 2.