Chapter 2 65
Test Descriptions
Gain Measurement Uncertainty
Figure 2-15 Typical Instrumentation Uncertainty for the range 30 to 70dB
Test Specification
Gain Measurement Uncertainty Range: -20 to >+40dB
Instrumentation Uncertainty: ±< 0.17dB
Related Adjustment
None