164 Appendix E
Caring for Connectors
Cleaning
Moisten the cloth with a small amount of alcohol and carefully insert it
into the connector to clean the interior surfaces. Use an illuminated
magnifying glass or microscope to see clearly the areas you wish to clean.
Drying Connectors
When you have cleaned a connector, always be sure that it is completely
dry before reassembling or using it. Blow the connector dry with a gentle
stream of clean compressed air and inspect it again under a magnifying
glass to be sure that no particles or alcohol residues remain.