56 Chapter 2
Test Descriptions
Noise Figure Range and Accuracy
Related Adjustment
None
Required Test Equipment
Table 2-12 Noise Figure Range and Accuracy test equipment
Equipment description Critical specification for equipment
substitution
Recommended model
Synthesized Sweeper Frequency Range: 50 MHz
Frequency Accuracy (CW): 0.02%
Power Level Range: -54dBm to -65dBm
83620/30/40/50B
Option 001 and 008
Attenuator/Switch Driver Compatible with 8494G/H Programmable step
attenuators
11713A
1 dB Step Attenuator Attenuation Range: 0 to 11dB
Frequency Range: 50 MHz
Connectors: 3.5mm (F)
Calibrated at 50 MHz with accuracy of 0.010 dB
8494G/H Option 002
Adapter 1 N8972/3A N Type (M) to 3.5mm (F)
N8974/5A 3.5mm Precision (F) to 3.5mm (F)
1250-1744
1250-1749 or 83059B
Adapter 83620/30 3.5mm Precision (F) to 3.5mm (F)
83640/50 2.4mm Precision (F) to 3.5mm (F)
1250-1749 or 83059B
11901B
RF Cable (X 2) 50 MHz 3.5mm (M) to 3.5mm (M), 61cm (24 in.) 11500E