10 Chapter 1
Preparing for Calibration and Performance Verification Test
Performance Verification and Adjustment Procedures
Performance Verification and Adjustment
Procedures
To perform a calibration run, the performance verification tests listed in
Table 1-4 below must be completed.
If any of the performance verification tests fail, perform the
corresponding calibration adjustment listed in Table 1-4.
If the performance verification tests fail and no calibration adjustment
procedure is available, the instrument must be returned to Agilent
Technologies for repair. Please contact your local Agilent Technologies
Customer Sales and Service Office for replacement parts or repair
service information. See the NFA User’s Guide for the address of your
nearest office.
Table 1-4 Verification and adjustment listings
Test no. Performance Verification Test Calibration Adjustments
1 10 MHz Out Frequency Reference Accuracy Test 2 (if applicable)
2 10 MHz Out Frequency Reference Adjustment
3 Input VSWR
4 Frequency Accuracy
5 Noise Source Supply Accuracy
6 Noise Figure Range and Accuracy
7 Gain Measurement Uncertainty
8 Instrument Noise Figure
9 Measurement Jitter

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight NFA Series and is the answer not in the manual?

Keysight NFA Series Specifications

General IconGeneral
BrandKeysight
ModelNFA Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals