32 Keysight P937xA PXIe Service Guide
Tests, Adjustments, and Troubleshooting
Non-Standards Compliant Verification
3-
Non-Standards Compliant Verification
To meet the criteria for non-standards compliant verification, perform the
preliminary checks and the performance tests while stopping to troubleshoot.
Refer to Figure 3-2 for test flow. Troubleshoot and repair the first problem
encountered without continuing to other tests. After you troubleshoot, repair,
and adjust, repeat the last failed portion and generate a new set of data.
Figure 3-2 Non-Standards Compliant Verification Flowchart