38 Keysight P937xA PXIe Service Guide
Tests, Adjustments, and Troubleshooting
System Verification
3-
Figure 3-4 NIST Traceability Path for Calibration and Verification Standard
Performing System Verification
The following verification procedure is automated by the network analyzer
firmware. The process for the verification is:
connect cables to the analyzer test ports
perform a calibration or recall a recent calibration
run the system verification program for the verification devices
Each time through the verification process, you are prompted to make
necessary connections and perform or recall a calibration as part of performing
the verification. If you select to perform a calibration, you are guided through