Keysight P937xA PXIe Service Guide 73
Removal and Replacement Procedures
Post-Repair Procedures
Post-Repair Procedures
After the replacement of an assembly, you must perform the service
procedures in the order listed in Table 5-1.
Table 5-1 Related Service Procedures
Replaced Assembly Adjustment, etc. TME Tests or Other Procedures
A1 Ingot Chassis No adjustment needed Perform the OF Ingot Utilities
Operator’s check
All TME based tests that apply
A2 Module Assembly No adjustment needed Perform the OF Ingot Utilities
Operator’s check
All TME based tests that apply
A3 Front Panel
Assembly
No adjustment needed Operator’s check
Connector No adjustment needed Operator’s check
All TME based tests that apply