Keysight P937xA PXIe Service Guide 49
Tests, Adjustments, and Troubleshooting
Adjustments
Source Adjustment
The source calibration is used to adjust your network analyzer for a flat source
power across its full frequency range. There are differences between each test
port; therefore, an adjustment is required for each port.
Equipment Used for the Source Adjustment
Procedure
1. Connect the USB power sensor and adapter to the analyzer Port 1. If using
a power meter, make the proper GPIB connections. Refer to “Get the
Hardware Required for your GPIB Instruments on page 27.
2. Press Utility > System > Service > Adjustments.
3. In the Select Desired Adjustment dialog box, click Source Adjustment.
4. Follow the instructions and prompts as they are displayed. If using a power
meter, ensure the GPIB settings for the sensor(s) are correct in the dialog
box.
Equipment Type Model or Part Number Alternate Model or Part Number
USB Thermocouple power sensor U8485A Option 200 U8485A power meter
Adapter, 3.5 mm -m- to 3.5 mm
-m-
83059A 85052-60014

Table of Contents

Other manuals for Keysight Streamline Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Streamline Series and is the answer not in the manual?

Keysight Streamline Series Specifications

General IconGeneral
BrandKeysight
ModelStreamline Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals