Electrical Requirements
4–4 SITE PREPARATION BEFORE INSTALLATION
Figure 4–1 Electrical Diagram of the X-ray Room and Connections
1 General mains 7bis X-ray warning lamp connecting terminal
2 Differential circuit breaker 8 Column connecting terminals
3 Red color actuator emergency stop push-button 9 X-ray remote control
4 Red color actuator emergency stop push-button 10 Door safety switch
5 Red warning lamp, power ON indicator 11 Mains outlet (for electric tools)
6 Unit mains connecting terminal 12 Ethernet outlet (RJ45/1)
7 Green warning lamp, ready state indicator 13 Contactor
N
L
G NL
N
L
LN
1A
12
10
9
7
11
5
4
3
2
1
6
8
7bis
RJ45
K9000C
13

Table of Contents