Electrical Requirements
KODAK 9000C and KODAK 9000C 3D Extraoral Imaging System Installation Guide (SM714)_Ed 01 4–5
A single-phase alternating current power supply is required.
The electrical installation specifications should be as follows:
Table 4–2 Electrical Installation Specifications
Supply Voltage
230V/240V
100V/110V/
130V
Frequency 50/60 Hz 50/60 Hz
Electrical supply 6 kW 6 kW
Line current
required
16A 20A
Cable cross-section
according to length
For 30 m:
2.5 mm2
For 10 m:
2.5 mm2
Max. line impedance 0.5 0.12
Differential circuit
breaker (2)
30 mA 30 mA
Maximum current to
trip circuit breaker
16A 20A Protect the power supply line with a
differential circuit breaker that trips at
maximum current.
Specifications of 2
red color actuator
emergency stop
push-button
(3) and (4)
250V 6A
UL listed
250V 6A
UL listed
Install these stop push-buttons to
simultaneously, switch off the
current to the active conductors of
the radiological installation and
exclude any other electric
equipment.
Locate (3) inside the x-ray room,
near the unit, for the operator to
quickly cut the power supply if
necessary.
Locate (4) outside the x-ray room,
near the x-ray remote control for
the operator to quickly cut the
power supply if necessary.
Maintain them in OFF (open)
position until a deliberate action is
performed.