Installation/Connection
<Out of Measurement Range>
[For xylv, T
∆∆
∆∆
uvLv, u'v'Lv or XYZ,Analyzer Mode]
When the measurement range is exceeded Digital display : “– – – – –”
Analog display : Not lit
LCD display : “OVER”
[For T
∆∆
∆∆
uvLv Mode]
T or uv are out of Digital display : “– – – – –”
the display range (T and uv)
[For Flicker Mode**]
When the measured value has Digital display : “– – – – –”
exceeded 999.9% Analog display : Not lit
LCD display : “FLICKER ERROR OVER”
When Lv(luminance) is the following Digital display : “– – – – –”
under 0.1cd/m
2
for LED Flicker Measuring ø27 Probe
Analog display : Not lit
(CA-P32/35)
LCD display :
FLICKER ERROR UNDER
under 0.3cd/m
2
for LED Flicker Measuring ø10 Probe
(CA-PS32/35)
**Flicker Mode is a function which can be used only when LED Flicker Measuring ø27 Probe(CA-P32/35) or LED Flicker Measuring ø10
Probe(CA-PS32/35) is connected.