Appendix J: Setting Up the Protector Echo Floor-Mount
Product Service 1-800-522-7658
176
Original instructions
Upper Cabinet Installation of the Protector
Echo Floor-Mounted Filtered Hood
Remove the upper cabinet assembly side panels and front panels prior to cabinet
placement. Then place the upper cabinet assembly on top of the lower base
assembly, being careful to clear the lower base assembly during placement. To
prevent spillage from seeping between the two liner assemblies, run a bead of
white RTV sealant between the sections once they have been properly aligned.
Mount the lower base to the upper cabinet using the ¼-20 hex head screws,
lockwashers, and nuts included in your hood package. See Figure J-2.
Figure J-1
Upper Cabinet Assy.
Lower
Base Assy.
Rear Vertical
Lower Sash with
inner wide weight
Front Vertical
Upper Sash with
Outer Weights
Apply bead of RTV
at inner joint
Lower
Perforated
Baffle
Lower Middle
Perforated Baffle
Typical Neutrodine
Filtration Technology

Table of Contents

Related product manuals