Chapter 3: Getting Started
Product Service 1-800-522-7658
15
Original instructions
All wiring for the filtered fume hood SHOULD be performed
by a licensed electrician and conform to all local codes.
Tout le câblage pour la hotte filtrée doit être effectuée par un
électricien agréé et conformément à tous les codes locaux.
The fluorescent light has been mounted inside the light module located on top of
the filtered hood. To change the fluorescent light bulbs in your filtered hood, you
must first remove the front panel from the hood. Next remove the filters, pre-
filters, and fan modules. Lift fixture up and replace any deflective bulbs. Reverse
order to reassemble.
Connecting the Service Lines to the
Protector Hood
The filtered hoods with service fixtures have been plumbed from the valve to the
hose connector or gooseneck for your installation convenience. Supply tubing
shall be provided by the qualified installer. Tubing can enter the filtered hood
from above, through the back, or through the work surface to make these
connections to the service fixtures.
NOTE: Inspect all fittings for leakage. Tighten the fittings
slightly if needed.
NOTE : Inspecter toutes les installations à la recherche de
fuite. Resserrer les installations légèrement si nécessaires.
CAUTION: Do not use oxygen with any standard service
fixture. Contact Labconco Customer Service for oxygen
fixture information.
PRUDENCE : Ne pas utiliser de l'oxygène avec l'accessoire de
service standard. Contacter le Service Clientèle de Labconco
pour les informations d'accessoire d'oxygène.
Should access to the filtered hood plumbing fixture bodies be required, remove
the service access plate on the hood front corner posts by loosening their
individual screws (see item 11, Figure A-1 in Appendix A). The valve body will
now be fully exposed for any service work that may be necessary. The service
fixtures supplied on your laboratory hood are designed for use with the following
services:
Air
Hot Water
Vacuum
Cold Water
Natural Gas See caution below
!
!

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Labconco Protector Echo 180 Series and is the answer not in the manual?

Labconco Protector Echo 180 Series Specifications

General IconGeneral
BrandLabconco
ModelProtector Echo 180 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals