Installation and Relocation 2-13
Caution: Do not run the pump until the chiller tubing is connected.
Running the pump without the chiller tubing may damage other
system components.
Figure 2.8 Tubing connections on the Echo system
7. Use the "cross-over" Ethernet cable to connect the Echo liquid handler
to the client PC. The client PC should have the Echo liquid handler
software already installed.
Fill the coupling fluid bottle and the fluidics circuit. See “Refill the coupling
fluid bottle” on page 6-8 and Cycle fluid through the Echo system” on
page 6-13.
Fluidics panel
From chiller (Water In)
To chiller (Water Out)
Compressed air line (Air In)
Vacuum line (Vacuum)