TOT AL CHLORINE
1. Press *IOK to turn the meter on.
1.3
2. Press *IOK to select Measure. Main Menu
* Measure
Data
Options
16:02:19 01/04/05
3. Scroll down and then press *IOK
to select Chlorine.
Measure
Turbidity
Color
* Chlorine
16:02:19 01/04/05
4. Scroll down and press *IOK to
select Test Total Chlorine.
Chlorine
Test Free
* Test Total
16:02:19 01/04/05
5. Rinse a clean tube (0290) with the
sample water. Fill the tube to the
10 mL line with the sample water.
Dry the tube with a lint-free cloth.
Put on a dry positioning ring. Cap
the tube.
52 CALIBRATION & ANALYSIS • CHLORINE