Calibration Procedure
The meter should be calibrated with platinum cobalt color standards. For the most
accurate results, a user calibration should be performed with LaMotte Color
Standards. The calibration should be done with a distilled or deionized water blank
and one color standard of known concentration. The concentration of the calibration
standard should be similar to the expected concentration of samples that will be
tested.
1. Press *IOK to turn the meter on.
1.3
2. Press *IOK to select Measure. Main Menu
* Measure
Data
Options
16:02:19 01/04/05
3. Scroll down and then press *IOK
to select Color.
Measure
Turbidity
* Color
Chlorine
16:02:19 01/04/05
4. Rinse a clean tube (0290) with
color-free (distilled or deionized)
water. Fill the tube to the 10 mL
line with color-free water. Dry the
tube with a lint-free cloth. Put on a
dry positioning ring. Cap the tube.
5. Open the meter lid. Insert the tube
into the chamber. Align the index
notch on the positioning ring with
the index arrow on the meter. Close
the lid and press *IOK to select
Scan Blank.
Color
* Scan Blank
16:02:19 01/04/05
CALIBRATION & ANALYSIS COL OR 71

Other manuals for LaMotte TC-3000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the LaMotte TC-3000 and is the answer not in the manual?

LaMotte TC-3000 Specifications

General IconGeneral
BrandLaMotte
ModelTC-3000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals