LaMotte TC-3000 - Liquid Chlorine Calibration Blanking; Adding Liquid Reagents for Calibration

LaMotte TC-3000
80 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4. Press *IOK to select Test Free
Chlorine.
Chlorine (L)
* Test Free
Test Total
16:02:19 01/04/05
5. Rinse a clean tube (0290) with the
sample water. Fill the tube to the
10 mL line with the sample water.
Dry the tube with a lint-free cloth.
Put on a dry positioning ring. Cap
the tube.
6. Open the meter lid. Insert the tube
into the chamber. Align the index
notch on the positioning ring with
the index arrow on the meter. Close
the lid and press *IOK to select
Scan Blank.
Free Chlorine (L)
* Scan Blank
16:02:19 01/04/05
7. Removethetubefromthemeter.
Add 5 drops of DPD 1A Free
Chlorine Reagent (P-6740) and
mix.
CALIBRATION & ANALYSIS CHLORINE 55
DPD
1A

Table of Contents