09/2021 ECO Gold 43 / 134
Contact with vapors from the heat transfer liquid
Use an extractor hood.
If possible, use a bath cover.
Bath overflow due to thermal expansion or
immersion of objects
Take the volume of external consumers into account.
Take into account the increase in volume with a rise in tem-
perature.
Hot vapor formation / discharge of boiling cooling-water on the cool-
ing coil
Filling of cooling coil with cooling water only admissible up to
T
max
of 100 °C!
Inadmissible operating temperatures;
temperature difference between outflow and product
too large
Property damage (consumers, external components)
Note that an externally controlled bath temperature, espe-
cially during a transient response, may differ substantially
from the set-point temperature.
Note the various limitation options (Tih, Til, T
max
, correction
limitation).
Set the overtemperature switch-off point T
max
according to
the heat transfer liquid. T
max
must be below the flash point.