13.6 Limiting the mains current consumption
If your mains fusing is below 16 A, the current consumption can be reduced in steps from 16 A to 8 A. The maximum
heating power is reduced correspondingly. Here, take into consideration whether other loads are connected to the same
fused circuit or whether your ECO thermostat is the only load.
Access to the main menu level is obtained by pressing the enter key
.
Selection and confirmation of Setup
Basic setup.
The adjacent menu window appears.
As well as Curr.Consumpt. the set value is displayed.
Select Curr.Consumpt. with .
Change the value with or .
Single figures can be selected by pressing or .
Confirm the input with the enter key .
-
By pressing (ESC) you are returned to the menu level without any change.
13.7 Entering the offset of the displayed temperature (calibration)
Deviations to the calibrated reference thermometers (e.g. LAUDA DigiCal) can be corrected internally by the "Off-
set" function.