09/2021 ECO Gold 99 / 134
16.3 Setpoint offset operating mode (Diff.set/actual)
It is possible to apply an offset value to the temperature, which is provided by an external temperature sensor and to
process it as the set value.
The bath temperature can therefore be operated, for example, -15 °C below the temperature of a reactor measured
by the external temperature sensor.
- Access to the main menu level is obtained by pressing the enter key
.
- Selection and confirmation of Setup Control
Setpoint offset.
The adjacent menu window appears.
As well as Diff.set/actual value the currently set offset value is displayed.
The adjacent menu window appears on selecting the menu point
Offset source.
- Select and confirm setpoint source with or and .
- The setpoint offset is deactivated with "off".
- Select and confirm Diff.set/actual value with or and .
The adjacent menu window appears. The minimum and maximum possi-
ble offset values and the current offset value are displayed.
- Change the value with or .
- Single figures can be selected by pressing or .
- By pressing (+/-) the arithmetic sign can be changed.
- Confirm your choice with the enter key .
-
By pressing (ESC) you are returned to the menu level without any change.