Lauda Ecoline RE 204 - Read Commands (Data Requested from Thermostat)

Lauda Ecoline RE 204
59 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Unit description
18.03.03
YACE0060
11
3.6.4 Read commands (data requested from thermostat)
Command Explanation
IN_PV_00 Read bath temperature (outflow temperature)
IN_SP_00 Read temperature setpoint
IN_SP_01 Read pump output step
IN_SP_02 Read operating mode of refrigerator
0 = OFF / 1 = ON 100% / 2 = ON 50% (reduced capacity RE 207, 210,
212, 220) / 3 = automatic operation
IN_SP_03 Read current overtemperature switch-off point
IN_PAR_00 Read current value of Xp
IN_PAR_01 Read current value of Tn
IN_MODE_00 Keys: 0 = free / 1 = inhibited
TYPE Read thermostat type
VERSION Read software version number
STATUS Read thermostat status 0 = OK, -1 = error
STAT Read error diagnosis response:
XXXXX ! X = 0 no error, X = 1 error
Char. 1 = Overtemperature
Char. 2 = low level
Char. 3 = pump blocked
Char. 4 = internal fault microcontroller 1
Char. 5 = internal fault microcontroller 2
For "_" use also " " (blank character).
The thermostat response is always in the fixed decimal format "XXX.XX" or for negative
values "-XXX.XX" or "ERR_X". (RS485 interface e.g. "A015_XXX.XX" or "A015_-XXX.XX"
or "A015_ERR_X").