Starting up
30
6.4.4.5 User calibration
Remove the external consumers and switch the setting knob of the pump to right
side. (Section 5.4).
A reference thermometer with necessary accuracy is required. Otherwise the factory
calibration should not be altered. The reference thermometer has to be inserted far
enough and long enough into the bath.
It is not allowed to calibrate to more than ± 3 °C. Multiple calibration to more than ± 3
°C cause internal faults (after 2 min ”E1006” or ”e16”).
The factory calibration will be lost through overwriting ! please work carefully!!!
and 3x
Directly from level 0 (setpoint selection), press the key
combination on the left, until the PARA-function is reached,
then
and 4x
switch to level 2 and move forward with key on the left..
>2,5 s
1. The display shows CAL and the current bath temperature.
To carry out a calibration, press the key longer than 2.5 s.
2. The actual value appears
or
3. Input the value indicated on the reference thermometer
with one to the two keys.
4. The additive calibration must be entered with the key
shown on the left.
5. Forward with key to "END", then